Konica Minolta CM-26 CM-25

Installing the Konica CM-25 & CM-26 driver

The Konica Minolta CM-25/26 driver is only available in Windows version.

Driver download for KM CM-25/26

The drivers are included in the Coraye plugin of the CM-26 which can be downloaded from the dashboard: https://dashboard.coraye.com/plugins

image-1629969396688.png

Once the plugin is downloaded, unzip it.

image-1629969427379.png

The driver is located in the folder: C: \ .... \ KM-CM26-Pilot-WIN-1.0.3 \ lib \ windows

When you connect the KM CM25 / 26 for the first time, it will appear as " USB Serial Device " in Windows Device Manager.

This means that the CM25 / 26 driver has never been installed.

image-1629969457391.png

To install the driver, right click on " USB serial device " then select " Update driver "

image-1629969482755.png

Select " Browse my computer ... " to indicate the path of the folder in which the driver is:
C: \ .... \ KM-CM26-Pilot-WIN-1.0.3 \ lib \ windows \ x64 \

image-1629969523891.png

image-1629969537488.png

When you have indicated the path, click on " Next "

image-1629969563633.png

The driver will install, and " USB Serial Device " will become " Measuring Instruments ".
This means that the driver is correctly installed.
Then click on the " 
Close " button

image-1629969600789.png


Modification of the port associated with the KM CM-25/26

To function properly, the serial port associated with the KM CM-25/26 must be COM4.

In the example below, the port assigned by Windows is COM3.

image-1629969669003.png

You will have to reassign the COM4 port.
Let's start by right-clicking on " 
Measuring Instruments ".
Then select " 
Properties "

image-1629969705176.png

Click on the " Advanced " button to access the advanced parameters for the COM3.
Modify the Port number and select COM4 from the list.
Then click on the " 
OK " button to validate the modification.

image-1629969742631.png

The COM4 port is assigned to the KM CM-25/26

image-1629969770356.png


Installation of the KM CM26 Pilot plugin

The installation procedure is described in the procedure: Installing and managing Plugins

To find out more, see the chapter : Installing and managing plugins

Since you have already downloaded the CM-26 plugin in the Dashboard, you can start with chapter 3 of the procedure


Using the KM CM-25/26 with Coraye

With the KM CM-25/26 you have the possibility to perform:
- The measurement of samples with the connected spectrophotometer
- The measurement of samples with the disconnected spectrophotometer
- The measurement of ranges to characterize prints on materials such as glass, metal,
etc ...

To measure samples with Coraye, launch the color capture module.

For more information, see chapter : Capturing a color

Add Color.png

Checking the connection

image-1629970652041.png

At this step you have the possibility:
- To download the measurements stored in the KM-CM 25/26 or / and
- To use the KM CM-25/26 to capture directly in Coraye, the color samples via a USB connection .


Import of measurements stored in the CM-26

By clicking on the " Import a reading " button, the samples stored in memory will be transferred to Coraye, then the spectrophotometer memory will be erased.

image-1629980772187.png


Measurements with the spectrophotometer connected via USB

Before clicking on the " Continue " button, remember to check the measurement mode you want to use (SCI or SCE) and also if you want to take into account the incidence of UVs.

UV mode selection

image-1629980833880.png

image-1629980857653.png

Selecting SCI or SCE mode

image-1629980882286.png

Coraye does not allow you to launch the calibration from its interface.
The calibration is therefore done on the KM CM-25/26 directly.

The procedure for calibrating the KM CM-25/26 is explained in chapter 2 of the manual, section "Calibration".
You will find the links to download these manuals at the end of the chapter.

When you have performed the calibration, click on " Instrument is ready " to proceed to sample measurement.

image-1629981046688.png

For the rest, refer to the chapter " Capturing a color "

For more information, see chapter: Capturing a color

Import of measurements taken

image-1629981151969.png


Creating an ICC profile with a KM SC-25/26

The method of creating a profile with a KM SC-25/26 is identical to other spectrophotometers.
All you have to do is follow the " Create a profile " procedure.

To find out more, see the chapter: Creating a profile from a standard test pattern

 

Do not forget to select the right reading modes
The measurement of the range being done patch by patch, it would be a shame to notice at the end of the measurement, that you made a mistake in choosing the SCI / SCE mode or UV100 / UV CUT400 especially after measuring 1000 patches ...


Measuring samples in a "Color Table"

The method for adding samples to a color table is the same as for other spectrophotometers.

For more information, see chapter: Color table

image-1629988953349.png


KM-25/26 User Manual

User manuals can be downloaded from the following addresses:
In English: 
https://www.konicaminolta.com/instruments/download/instruction_manual/color/pdf/cm-26dg_instruction_eng.pdf

In French: https://www.konicaminolta.com/instruments/download/instruction_manual/color/pdf/cm-26dg_instruction_fra.pdf

In German: https://www.konicaminolta.com/instruments/download/instruction_manual/color/pdf/cm-26dg_instruction_deu.pdf Other languages: https://www.konicaminolta.com/instruments/download/instruction_manual/color/index .html


SCI / SCE mode with or without UV

SCI Mode: Specular Component Included

SCE Mode: Specular Component Excluded

Link to a Konica Minolta blog explaining the difference between SCI and SCE mode: 
https://sensing.konicaminolta.us/blog/specular-component-included-sci-vs-specular-component- excluded-sce / Mode UV 100: Measurement with UV source to react optical brighteners

CUT400 UV mode: Measurement with filtered UV below 400 nm to ignore optical brighteners

Example of measurement carried out on a glossy Chromaluxe plate

A gray sample was measured with the four reading combinations: The display of the reflectance curves makes it possible to visualize the difference in measurement between the different modes:
- SCI - UV100 (Red)
- SCE - UV100 (Blue)
- SCI - UV CUT400 (Orange)
- SCE - UV CUT400 (Cyan)

image-1629989259107.png

Note: On the right column you can directly obtain the information associated with the selected sample.