Compare and edit chart measures

Before taking the average of several chart measurements, intended for the creation of an icc characterization profile, it is advisable to check the measurements and correct them if necessary.


In this tutorial, we will measure 3 identical ECI2002R charts. The goal is to check if there is not too much difference between these charts and to check if measurement errors have not crept in during the reading.


Import of measurements made in the " Create profile " module

To find out more, see the chapter: Create an icc profile from an average of measured charts

image-1629206215413.png

image-1629992484342.png

The imported measurements will appear in the left column


Creation of a reference file

Click on the " Create a reference " button

image-1629992558107.png

The ECI2002R test charts being in CMJN mode, we will click on the "CMJN" button

image-1629992593745.png

We will choose the "ECI2002R sample A" measurement as a reference to compare it with the other two.

Drag the "ECI2002R sample A" measurement into the " Drag the color table here " area

image-1629992626957.png

Then click on the button " Use the Labs of the standard "

image-1629992654778.png

Click on the " Next " button located at the bottom of the test chart

image-1629992682094.png

Name the reference file then click on " Save and exit "

image-1629992706566.png

The reference file appears in the left column.

image-1629992736588.png

Make sure the spectrophotometer is plugged in then right click on the reference file then select " Start control "

image-1629992763120.png

Select the options the options of the spectrophotometer then click on "I'm ready"

image-1629992788360.png

Position the spectrophotometer in its calibration position
image-1629199850021.png
Then click on the " Start calibration " button
image-1629992880706.png

Comparison of a measurement against the reference measurement

Drag the "ECI2002 sample B" measurement file onto the "Reading # 1" button

image-1629992926914.png

When the measurement is loaded, click on the " Measurement finished " button

image-1629992956218.png

The report is displayed and indicates that there are samples that have an unusually high difference.
This may be due to a defective target or to a reading error.

image-1629992990033.png

In the "Reading report" area we will identify the bad measurements.
Two measurements are problematic:
- The 18th measurement in column 13 => M10
- The 10th measurement in column 14 => N18 We are going to measure them again.

image-1629993040055.png

When the measurement is loaded, the color hover displays the deltaE values.
Click on the color to re-measure.

image-1629993095341.png

The spectrophotometer will ask you for a calibration then measure the sample on the printed test chart.

I1 Pro2 measure.png

Check the new value after measurement.

image-1629993275654.png

Perform the same operation for the second value.
Check the new value after measurement. When the measurements are finished, click on the " Measurement finished " button

image-1629993373851.png

The new report is displayed

image-1629994061187.png

Carry out the same operation with the other measurement "ECI2002R sample C"

The average of the verified measurements can be used to create the icc profile

image-1629993533931.png

To find out more, see the chapter:  Create an icc profile from an average of measured charts